• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. RUBY COIL; HCG, KRD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. RUBY COIL; HCG, KRD Back to Search Results
Catalog Number RBY4C2060
Medical Device Problem Code Mechanical Jam (2983)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/10/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report numbers: 3005168196-2018-01758, 3005168196-2018-01759, 3005168196-2018-01760.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure in the splenic artery using ruby coils and a lantern delivery microcatheter (lantern).During the procedure, the physician successfully deployed and detached a ruby coil in the target vessel using a lantern.While attempting to advance a new ruby coil through the lantern, the ruby coil became stuck; therefore, it was removed.It was reported that the same issue occurred with two other ruby coils; therefore, the lantern and ruby coils were removed.The procedure was completed using a new lantern and other coils.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RUBY COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key7857461
Report Number3005168196-2018-01757
Device Sequence Number1981069
Product Code HCG
UDI-Device Identifier00814548013251
UDI-Public00814548013251
Combination Product (Y/N)Y
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K120330
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/10/2018
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/05/2025
Device Catalogue NumberRBY4C2060
Device Lot NumberF75465
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/10/2018
Initial Report FDA Received Date09/07/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age53 YR
Date Report Sent to FDA01/01/2005
-
-