• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD ACCESS SYSTEMS POWER-TRIALYSIS SHORT-TERM STRAIGHT DIALYSIS CATHETER FULL PROCEDURE TRAY 13 F; CATHETER, HEMODIALYSIS, TRIPLE LUMEN, NON-IMPLANTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS POWER-TRIALYSIS SHORT-TERM STRAIGHT DIALYSIS CATHETER FULL PROCEDURE TRAY 13 F; CATHETER, HEMODIALYSIS, TRIPLE LUMEN, NON-IMPLANTED Back to Search Results
Model Number 5605150
Health Effect - Clinical Code Air Embolism (1697)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device has not been returned to the manufacturer for evaluation.A lot history review (lhr) of recs3172 showed three other similar product complaint(s) from this lot number.The complaints for this lot number (recs3172) have been reported from the same facility.
 
Event or Problem Description
The facility reported an event in which a patient being treated with tl acute power dialysis catheter coded.When an ultrasound was performed, air was discovered in both the atrium and the ventricle.All of the department was notified to pay close attention to the tubing, machine, and catheter, and they were able to rule out the dialysis machine as the problem.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWER-TRIALYSIS SHORT-TERM STRAIGHT DIALYSIS CATHETER FULL PROCEDURE TRAY 13 F
Common Device Name
CATHETER, HEMODIALYSIS, TRIPLE LUMEN, NON-IMPLANTED
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key7857927
Report Number3006260740-2018-02408
Device Sequence Number16629463
Product Code NIE
UDI-Device Identifier00801741086953
UDI-Public(01)00801741086953
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K133456
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number5605150
Device Catalogue Number5605150
Device Lot NumberRECS3172
Was Device Available for Evaluation? No
Event Location Hospital
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date09/07/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-