• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523LNAH; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523LNAH; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-523LNAH
Medical Device Problem Code Excess Flow or Over-Infusion (1311)
Health Effect - Clinical Code Hypoglycemia (1912)
Date of Event 08/15/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
The customer reported via phone call that they experienced low blood glucose as well as insulin pump had a over delivering alarm.Blood glucose reading was 25 mg/dl at the time of incident.Current blood glucose reading was 98 mg/dl.Customer treated their blood glucose with food and glucose tablets.The customer alleged the pump was over-delivering insulin.Troubleshooting did not find any issues with the insulin pump.The insulin pump will be returned for analysis.
 
Additional Manufacturer Narrative
No unexpected no delivery alarm noted during testing.Device received with all operating currents within specification and passed functional testing including the rewind test, self-test,error test, basic occlusion test, occlusion test, prime test, excessive no delivery test and displacement test.Also, unit passed the delivery accuracy test at 0.08830 inches.No possible over delivery anomaly noted.The information provided in concomitant reporting section with the initial report was incorrect.The correct information has been included with this report.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523LNAH
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key7858884
Report Number2032227-2018-25187
Device Sequence Number19166825
Product Code OYC
UDI-Device Identifier00643169151284
UDI-Public(01)00643169151284
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-523LNAH
Device Catalogue NumberMMT-523LNAH
Device Lot NumberA2523LNAHJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/07/2018
Initial Date Received by Manufacturer 08/29/2018
Supplement Date Received by Manufacturer12/05/2018
Initial Report FDA Received Date09/09/2018
Supplement Report FDA Received Date12/06/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age68 YR
Patient Weight120
-
-