MAUDE Adverse Event Report: WARSAW ORTHOPEDICS CD HORIZON® SPINAL SYSTEM; APPLIANCE, FIXATION, SPINAL INTERLAMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WARSAW ORTHOPEDICS CD HORIZON® SPINAL SYSTEM; APPLIANCE, FIXATION, SPINAL INTERLAMINAL Back to Search Results
Catalog Number 7576545
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 08/13/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product was not returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.Due to imdrf harmonization, some previously submitted device, method, result, and conclusion codes related to this event may have been updated.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient underwent tlif (transforaminal lumbar interbody fusion) surgery.Post-operatively, the head of the screw seperated from the screw body.No patient complications were reported.
 
Additional Manufacturer Narrative
Product analysis result: visual macroscopic dimensional visual inspection confirms the bone screw disassembled from the head.The ring and crown were still assembled in the head.About half of the ring is missing and the other portion is still seated in the groove of the head.The broken portion of the ring is missing and not returned.Macroscopic inspection of the inner head surface of the screw revealed multiple witness marks consistent with the threads of the bone screw head.Witness marks on the bone screw head are noted consistent with where the ring made contact with the bone screw head before the ring broke.The above observations are consistent with overload as the mechanism of failure.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD HORIZON® SPINAL SYSTEM
Common Device Name
APPLIANCE, FIXATION, SPINAL INTERLAMINAL
Manufacturer (Section D)
WARSAW ORTHOPEDICS
2500 silveus crossing
warsaw IN 46582
MDR Report Key7859014
Report Number1030489-2018-01219
Device Sequence Number1431672
Product Code KWP
UDI-Device Identifier00885074031156
UDI-Public00885074031156
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K032265
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/24/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number7576545
Device Lot Number0636480W
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/09/2019
Initial Date Received by Manufacturer 08/13/2018
Supplement Date Received by Manufacturer05/09/2019
Initial Report FDA Received Date09/10/2018
Supplement Report FDA Received Date05/24/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-