DAVIS & GECK CARIBE LTD BIOSYN; SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID Back to Search Results
Model Number SM-632
Medical Device Problem Code Unraveled Material (1664)
Health Effect - Clinical Codes Wound Dehiscence (1154); Seroma (2069); No Code Available (3191)
Date of Event 08/21/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during a procedure, on ovarian pedicle uterine pedicle linear and sub quticular, the patient came back in and needed another surgery due to drainage, seroma and wound opening.They re-opened and re-sutured as a surgical intervention.The incision was extended due to the device problem and there would be an additional operation to correct the issue.The surgical time was extended for more than thirty minutes.The patient¿s hospitalization was extended to ten days.
 
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of four representative samples of a device.Visual inspection and functional testing confirmed there were no abnormalities that would have caused or contributed to the reported condition.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOSYN
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DO  0101
MDR Report Key7859953
Report Number9612501-2018-01614
Device Sequence Number16635714
Product Code GAM
UDI-Device Identifier10884521033160
UDI-Public10884521033160
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K000037
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/15/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2023
Device Model NumberSM-632
Device Catalogue NumberSM-632
Device Lot NumberD8C1761FX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/04/2018
Initial Date Received by Manufacturer 08/21/2018
Supplement Date Received by Manufacturer01/23/2019
Initial Report FDA Received Date09/10/2018
Supplement Report FDA Received Date02/15/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age6 MO
Patient Weight1
-
-