Model Number 8253001
Medical Device Problem Code Failure to Sense (1559)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/16/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Other relevant device: product id: 8253200.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
A health care provider (hcp) reported via a manufacturer representative that while doing the electrode check there were x's during the thyroidectomy procedure.The site switched out the patient interface and it ended up working; however, when they were stimulating during the procedure they all of a sudden lost stimulation.The manufacturer representative tested out the system in another room and was getting a response on all the channels.There was no delay in the procedure.There was no patient impact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NIM® 3.0 MAINFRAME
Common Device Name
STIMULATOR, NERVE
Manufacturer (Section D)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
MDR Report Key7860136
Report Number1045254-2018-00421
Device Sequence Number13949885
Product Code ETN
UDI-Device Identifier00643169547049
UDI-Public00643169547049
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K083124
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number8253001
Device Catalogue Number8253001
Device Lot Number211862738
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/16/2018
Initial Report FDA Received Date09/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-