| Catalog Number |
0006412 |
| Medical Device Problem Code |
False Positive Result (1227)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/02/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The immucor laboratory tested retention product on 22aug2018, which performed as expected.The immucor laboratory also received and tested a blood sample on 22aug2018 from the customer site ((b)(4)) and obtained expected negative outcomes.This did not reproduce the customer result outcomes from (b)(6) 2018, but did reproduce the customer result outcomes from more recent blood samples.(b)(4).
|
| |
|
Event or Problem Description
|
|
On 17aug2018, a customer site reported an unexpectedly positive outcome with anti-d (series 4) monoclonal blend by manual test tube method, when tested on (b)(6) 2018.
|
| |
|