• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER NEUROVASCULAR CORK NEUROFORM ATLAS STENT SYSTEM 4.0 X 24MM WITHOUT TIP; STENT, INTRACRANIAL NEUROVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER NEUROVASCULAR CORK NEUROFORM ATLAS STENT SYSTEM 4.0 X 24MM WITHOUT TIP; STENT, INTRACRANIAL NEUROVASCULAR Back to Search Results
Catalog Number M003SZAS40240
Medical Device Problem Code Activation Failure (3270)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/24/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device implanted.
 
Event or Problem Description
It was reported that during the procedure, the stent was deployed from the right posterior cerebral artery (pca) to the basilar artery (ba).The stent opened proximally in the basilar artery; however, the distal of the stent did not open in the posterior cerebral artery.There were no clinical consequences to the patient reported.
 
Additional Manufacturer Narrative
The device history record (dhr) review confirms that the device met all material, assembly and performance specifications.The subject device was not returned for analysis; therefore, visual, dimension and functional testing could not be performed.The device history record (dhr) review confirms that the device met all material, assembly and performance specifications.The subject device was not returned for analysis; therefore, visual, dimension and functional testing could not be performed.As per the additional information provided in the complaint, the device was inspected and confirmed to be in a good condition during/after unpacking and preparation, and the device was prepared as per direction for use (dfu).In case of this complaint, there are several potential causes for the reported issue, but because review of available information failed to identify a definitive cause and since the product was not returned for analysis, a cause of undetermined was assigned to the as reported issue 'nv ¿ stent failed/unable to open'.
 
Event or Problem Description
It was reported that during the procedure, the stent was deployed from the right posterior cerebral artery (pca) to the basilar artery (ba).The stent opened proximally in the basilar artery; however, the distal of the stent did not open in the posterior cerebral artery.There were no clinical consequences to the patient reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEUROFORM ATLAS STENT SYSTEM 4.0 X 24MM WITHOUT TIP
Common Device Name
STENT, INTRACRANIAL NEUROVASCULAR
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
MDR Report Key7860479
Report Number3008881809-2018-00394
Device Sequence Number1337366
Product Code NJE
UDI-Device Identifier07613252656203
UDI-Public07613252656203
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
H020002/S046
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/07/2023
Device Catalogue NumberM003SZAS40240
Device Lot Number20381712
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/24/2018
Supplement Date Received by Manufacturer10/02/2018
Initial Report FDA Received Date09/10/2018
Supplement Report FDA Received Date10/24/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-