• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIODYNAMICS ANGIODYNAMICS / SMART PORT; PORT & CATHETER, IMPLANTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIODYNAMICS ANGIODYNAMICS / SMART PORT; PORT & CATHETER, IMPLANTED Back to Search Results
Catalog Number H787CT75STSA0
Medical Device Problem Codes Crack (1135); Difficult to Remove (1528)
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 07/19/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the device history records was performed for the indicated packaging and component lots for any deviations related to the reported defect of the complaint.The review confirms that the lots met all material, assembly and performance specifications.The recent angiodynamics complaint report was reviewed for the smartport product family and the failure mode "catheter difficult removal." no adverse trend was indicated.The root cause of the reported event was unable to be determined as no sample was returned for evaluation.Port placement, patient anatomy and drugs used during treatment are potential contributing factors of the difficult to remove catheter tubing.The directions for use packaged with the smart port contain guidance on port placement, maintenance and explantation.(b)(4).
 
Event or Problem Description
As reported by angiodynamics' distributor in (b)(4), a smart port had been implanted in a pediatric patient (age unknown) on (b)(6) 2015 for treatment of acute lymphoblastic leukemia.Difficulty was encountered when explanting the port/catheter on (b)(6) 2018: the catheter was stuck, and when pulled on, broken.The catheter was successfully retrieved.The used device will not be returned to angiodynamics.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANGIODYNAMICS / SMART PORT
Common Device Name
PORT & CATHETER, IMPLANTED
Manufacturer (Section D)
ANGIODYNAMICS
10 glens falls technical park
glens falls NY 12801
MDR Report Key7860588
Report Number1317056-2018-00171
Device Sequence Number11254530
Product Code LJT
UDI-Device IdentifierH787CT75STSA0
UDI-PublicH787CT75STSA0
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K062414
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2018
Reporter Type Manufacturer
Report Source distributor,foreign
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/30/2015
Device Catalogue NumberH787CT75STSA0
Device Lot Number4824929
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/14/2018
Initial Report FDA Received Date09/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-