Model Number 1328
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 08/23/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This device is used for treatment not diagnosis.The ventricular assist system is indicated for use as a bridge to cardiac transplantation and destination therapy in patients who are at risk of death from refractory end-stage left ventricular heart failure.The system is designed for in-hospital and out-of-hospital settings.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the apical coring tool felt ¿stiff¿ during the implant procedure.It was noted to not have a smooth motion when deployed, the surgeon continued to use it for the implantation of the ventricular assist device (vad).The apex core tool was discarded.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Product event summary: one (1) apical coring tool associated with surgical tool kit was not returned for analysis.Review of the manufacturing documentation confirmed that the associated tool met all requirements for release.The reported event could not be confirmed since the device was not returned and there is no evidence or supporting data provided.There is no evidence to suggest that a component malfunction caused or contributed to the reported event.Based on the limited information available, possible causes of the reported event can be attributed to multiple factors, including but not limited to higher wall thickness of the left ventricle and/or user perception.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTWARE VENTRICULAR ASSIST SYSTEM- APEX CORING
Common Device Name
VENTRICULAR (ASSISST) BYPASS
Manufacturer (Section D)
HEARTWARE, INC.
14400 nw 60th ave
miami lakes FL 33014
MDR Report Key7860667
Report Number3007042319-2018-04313
Device Sequence Number1633360
Product Code DSQ
UDI-Device Identifier00888707000451
UDI-Public00888707000451
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
P100047
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/31/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2020
Device Model Number1328
Device Catalogue Number1328
Device Lot Number1451010
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/23/2018
Supplement Date Received by Manufacturer01/30/2019
Initial Report FDA Received Date09/10/2018
Supplement Report FDA Received Date01/31/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MCS UNKNOWN VAD
Patient Age72 YR
-
-