Model Number 2426-0007 |
Device Problems
Fluid/Blood Leak (1250); Material Integrity Problem (2978)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The affected product has been received and the evaluation is pending.A follow up report will be submitted once the evaluation is completed.
|
|
Event Description
|
The customer reported that last weekend, tubing leaked from a hole that developed above the section that inserts into the pump module during patient use.There was no patient harm.
|
|
Manufacturer Narrative
|
The customer¿s report of the set leaking was confirmed and replicated; however, the leak was below the lower fitment, not above the upper fitment as reported.Visual inspection showed that the vinyl tubing had a 0.0346 inch long slice cut approximately 11 1/2 inches below the lower fitment.Functional and pressure testing was performed and a leak was observed from the cut below the lower fitment.The cause of the leak was a cut in the tubing.The slice cut is consistent with damage of a box cutter being used to open the package or a sharp object being used at some point during infusion to remove tape from the tubing.
|
|
Event Description
|
The customer reported that last weekend, tubing leaked from a hole that developed above the section that inserts into the pump module during patient use.There was no patient harm.
|
|
Search Alerts/Recalls
|