• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TRIVIDIA HEALTH, INC. TRUEMETRIX; BLOOD GLUCOSE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TRIVIDIA HEALTH, INC. TRUEMETRIX; BLOOD GLUCOSE SYSTEM Back to Search Results
Model Number TRUEMETRIX
Medical Device Problem Code Unable to Obtain Readings (1516)
Health Effect - Clinical Codes Headache (1880); Dizziness (2194)
Date of Event 08/15/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Returned meter evaluated with defect found.Returned test strips evaluated with no defect found.Qc investigation on 9/4/2018 returned meter produce e-7 error code.Final root cause: miscellaneous user induced contamination on the strip connector test strip udi# (b)(4).Manufacturer contacted customer on (b)(6) 2018 in a follow-up call in order to ensure the customer's condition since the initial call; customer condition improved.No symptoms or medical attention reported since the original call.Customer satisfied with the replacement product.
 
Event or Problem Description
Consumer reported complaint for e-3 blood glucose results.Wife is calling on behalf of the customer.The expected fasting blood glucose test result range is undisclosed.The customer did report symptoms of having a headache and feeling dizzy.Medical attention is not reported as a result of the actual blood glucose results and reported symptoms.The product storage location is undisclosed.During the call on (b)(6) 2018, a blood test was performed by the customer and produced test results of e-3 using true metrix meter.The test strip lot manufacturer's expiration date is 10/20/2019 and open vial date is undisclosed.The meter memory was not reviewed for previous test result history.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUEMETRIX
Common Device Name
BLOOD GLUCOSE SYSTEM
Manufacturer (Section D)
TRIVIDIA HEALTH, INC.
2400 nw 55th court
fort lauderdale FL 33309
MDR Report Key7861361
Report Number1000113657-2018-00936
Device Sequence Number13949896
Product Code NBW
UDI-Device Identifier00021292007041
UDI-Public(01)00021292007041
Combination Product (Y/N)N
Initial Reporter StateVA
PMA/510(K) Number
K140100
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial
Report Date (Section B) 09/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberTRUEMETRIX
Device Catalogue NumberRE4H01-81
Device Lot NumberMV2863
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/23/2018
Was the Report Sent to FDA? No
Event Location Home
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/15/2018
Initial Report FDA Received Date09/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-