| Model Number |
TRUEMETRIX |
| Medical Device Problem Code |
Unable to Obtain Readings (1516)
|
| Health Effect - Clinical Codes |
Headache (1880); Dizziness (2194)
|
| Date of Event |
08/15/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Returned meter evaluated with defect found.Returned test strips evaluated with no defect found.Qc investigation on 9/4/2018 returned meter produce e-7 error code.Final root cause: miscellaneous user induced contamination on the strip connector test strip udi# (b)(4).Manufacturer contacted customer on (b)(6) 2018 in a follow-up call in order to ensure the customer's condition since the initial call; customer condition improved.No symptoms or medical attention reported since the original call.Customer satisfied with the replacement product.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for e-3 blood glucose results.Wife is calling on behalf of the customer.The expected fasting blood glucose test result range is undisclosed.The customer did report symptoms of having a headache and feeling dizzy.Medical attention is not reported as a result of the actual blood glucose results and reported symptoms.The product storage location is undisclosed.During the call on (b)(6) 2018, a blood test was performed by the customer and produced test results of e-3 using true metrix meter.The test strip lot manufacturer's expiration date is 10/20/2019 and open vial date is undisclosed.The meter memory was not reviewed for previous test result history.
|
| |
|