W.L. GORE & ASSOCIATES AORTIC EXCLUDER AAA ENDOPROSTHESIS (LOW PERM); SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Catalog Number PXT261412
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 08/14/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Concomitant medical products: additional components implanted on (b)(6) 2007 - pxc201200/05072934 (b)(4), pxc181400/05225188 (b)(4).
 
Event or Problem Description
On (b)(6) 2007, this patient underwent endovascular repair of an abdominal aortic aneurysm using gore® excluder® aaa endoprosthesis.At the time of repair the aneurysm measured 6.5 cm.The patient was then lost to follow up.On (b)(6) 2018, the patient had a follow up examination where it was revealed that the aneurysm measured 11.5 cm.Non-contrast computed tomography showed what appeared to be a proximal type i endoleak at the proximal neck, and distal type i endoleaks from both common iliac arteries.A reintervention is planned where the physician stated he will coil the left hypogastric artery, then extend both iliac arteries, as well as snorkel the left renal artery and place a cook aortic cuff at the proximal neck.
 
Additional Manufacturer Narrative
Manufacturer narrative/corrected data contact office name/address (and manufacturing site for devices)name and address of manufacturer's contact office.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AORTIC EXCLUDER AAA ENDOPROSTHESIS (LOW PERM)
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
W.L. GORE & ASSOCIATES
flagstaff AZ
MDR Report Key7861598
Report Number3007284313-2018-00254
Device Sequence Number19328136
Product Code MIH
UDI-Device Identifier00733132610211
UDI-Public00733132610211
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
P020004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2007
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/31/2010
Device Catalogue NumberPXT261412
Device Lot Number05170681
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date09/10/2018
Supplement Report FDA Received Date11/08/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-