The customer reported the sonde (catheter) plug had a loose connection and as a result the patient had to be resuscitated.With these sondes a couple problems occur that threaten the life of the patient.In addition there is a loose connection at the plug, so there is no stimulation.This is what happened today in the imc.The patient came with the av block iii°, came directly from the emergency department to the cath lab, immediately received the temporary pacemaker and in spite of the correct position of the sonde (catheter) the patient had to be resuscitated in the imc and an external stimulation by a defibrillator was necessary.It turned out that there was a loose connection of the plug.
|
|
Device used in treatment.The device was not returned for analysis, therefore, the clinical observation could not be confirmed.The investigation will focus on a review of product documentation.A review of device history record (dhr) was conducted and there were no manufacturing rejects or anomalies recorded in the dhr affecting released product.The released device passed all in-process and qa final inspection steps before shipping to the customer.Capa was previously opened to investigate the root cause and to implement corrective action to prevent recurrence of this issue.The event will be re-evaluated if additional information becomes available.Oscor will continue to monitor this event type and risk.Field action has been initiated for tb lead.The tb leads are inspected visually and tested for electrical continuity 100% by qa prior to shipping to the customer per procedure temporary lead final inspection (model tb).Per instructions for use tb without needle/cannual (ifu) precautions are provided to the user: do not connect to a/c power.Inappropriate connection of exposed pin(s) to an ac power source may pose serious risk of injury or death.Avoid kinking or bending the lead to minimize the risk of damage to internal wires.Avoid subjecting the device to unusual stresses.The device should be stored at temperatures between 5° c (41° f) and 30° c (86° f).Oscor inc.Is submitting the above report to comply with 21 cfr part 803, the medical device reporting regulation.The report is based upon information obtained by oscor inc.Which the company may not have been able to investigate or verify prior to the date the report was required by fda.This report does not constitute an admission that the device, oscor inc., or its employees, caused or contributed to the event described in this report.Nor does this report reflect a conclusion by fda, oscor inc., or its employees, that the report constitutes an admission that the device, oscor inc., or its employees caused or contributed to the event described in this report.
|