| Catalog Number |
75HA70 |
| Medical Device Problem Code |
Inflation Problem (1310)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/10/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluation in progress.
|
| |
|
Event or Problem Description
|
|
It was reported that during a pre-test check, the cuff component of the tracheostomy tube failed to inflate equally.No death or serious injury was reported in connection with this incident.
|
| |
|
Additional Manufacturer Narrative
|
|
One mid-range adjustable hyperflex tracheostomy (trach) was returned for analysis in good condition.Inflation test was performed by using a syringe to inflate the cuff with 20cc of air.The cuff symmetry did not exceed the 1/3:2/3 for an aire cuff trach.The cuff was removed from the shaft and four thickness measurements of the cuff were taken using a digital mitutoyo thickness gage; all measurements are within specification.Based on the evidence, the root cause is unknown as the complaint could not be confirmed.
|
| |
|
Additional Manufacturer Narrative
|
|
One mid-range adjustable hyperflex tracheostomy (trach) was returned for analysis in good condition.Inflation test was performed by using a syringe to inflate the cuff with 20cc of air.The cuff symmetry did not exceed the 1/3:2/3 for an aire cuff trach.The cuff was removed from the shaft and four thickness measurements of the cuff were taken using a digital mitutoyo thickness gage; all measurements are within specification.Based on the evidence, the root cause is unknown as the complaint could not be confirmed.
|
| |
|