• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. AIRE-CUF TRACHY TUBE 7.0MM ADLT MID-RNGE,ADJUST HYPERFLEX; TUBE TRACHEOSTOMY AND TUBE CUFF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. AIRE-CUF TRACHY TUBE 7.0MM ADLT MID-RNGE,ADJUST HYPERFLEX; TUBE TRACHEOSTOMY AND TUBE CUFF Back to Search Results
Catalog Number 75HA70
Medical Device Problem Code Inflation Problem (1310)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/10/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device evaluation in progress.
 
Event or Problem Description
It was reported that during a pre-test check, the cuff component of the tracheostomy tube failed to inflate equally.No death or serious injury was reported in connection with this incident.
 
Additional Manufacturer Narrative
One mid-range adjustable hyperflex tracheostomy (trach) was returned for analysis in good condition.Inflation test was performed by using a syringe to inflate the cuff with 20cc of air.The cuff symmetry did not exceed the 1/3:2/3 for an aire cuff trach.The cuff was removed from the shaft and four thickness measurements of the cuff were taken using a digital mitutoyo thickness gage; all measurements are within specification.Based on the evidence, the root cause is unknown as the complaint could not be confirmed.
 
Additional Manufacturer Narrative
One mid-range adjustable hyperflex tracheostomy (trach) was returned for analysis in good condition.Inflation test was performed by using a syringe to inflate the cuff with 20cc of air.The cuff symmetry did not exceed the 1/3:2/3 for an aire cuff trach.The cuff was removed from the shaft and four thickness measurements of the cuff were taken using a digital mitutoyo thickness gage; all measurements are within specification.Based on the evidence, the root cause is unknown as the complaint could not be confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIRE-CUF TRACHY TUBE 7.0MM ADLT MID-RNGE,ADJUST HYPERFLEX
Common Device Name
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
MDR Report Key7861776
Report Number3012307300-2018-03586
Device Sequence Number1337409
Product Code JOH
UDI-Device Identifier15021312013849
UDI-Public15021312013849
Combination Product (Y/N)N
Initial Reporter CountryTC
PMA/510(K) Number
K081440
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/13/2022
Device Catalogue Number75HA70
Device Lot Number3429657
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/06/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/12/2018
Supplement Date Received by Manufacturer09/20/2018
Initial Report FDA Received Date09/10/2018
Supplement Report FDA Received Date10/18/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-