MAUDE Adverse Event Report: BIOMERIEUX INC. VITEK® 2 AST-GN69 TEST KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMERIEUX INC. VITEK® 2 AST-GN69 TEST KIT Back to Search Results
Catalog Number 413400
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Patient Involvement (2645)
Type of Reportable Event Malfunction
Event or Problem Description
A customer in the united states notified biomérieux of false susceptible results for an escherichia coli survey strain when testing with the vitek® 2 ast-gn69 test kit.The customer obtained a susceptible result (mic=8) for ampicillin/sulbactam.The expected result for this survey strain is resistant.The customer was not able to repeat testing because they no longer had the isolate available.As there is no patient associated with this quality control strain, there is no adverse event related to any patient's state of health.A biomérieux internal investigation will be initiated.
 
Additional Manufacturer Narrative
The customer no longer has their api survey strain for this investigation.Submittal of the isolate is required in order to confirm a vitek 2 discrepancy compared to the reference method.Vitek 2 ast-gn69 lot# 5890624403 met final qc release criteria and passed qc performance testing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VITEK® 2 AST-GN69 TEST KIT
Common Device Name
VITEK® 2 AST-GN69 TEST KIT
Manufacturer (Section D)
BIOMERIEUX INC.
595 anglum road
saint louis MO 63042
MDR Report Key7861873
Report Number1950204-2018-00348
Device Sequence Number19193976
Product Code LON
UDI-Device Identifier03573026398460
UDI-Public03573026398460
Combination Product (Y/N)N
Initial Reporter StateAR
Initial Reporter CountryUS
PMA/510(K) Number
N50510: S 97
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/12/2019
Device Catalogue Number413400
Device Lot Number5890624403
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/14/2018
Supplement Date Received by Manufacturer10/30/2018
Initial Report FDA Received Date09/10/2018
Supplement Report FDA Received Date11/28/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-