BOSTON SCIENTIFIC - GALWAY SYNERGY II EVEROLIMUS-ELUTING PLATINUM CHROMIUM CORONARY STENT SYSTEM; BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM Back to Search Results
Model Number H7493926020350
Device Problem Defective Component (2292)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Device is a combination product.(b)(4).Device evaluated by mfr: the synergy ii us mr 3.50 x 20 mm stent delivery system (sds) was returned for analysis.A visual examination of the stent found that the first proximal stent row was damaged with stent struts lifted and pulled distally.The undamaged section stent outer diameter was measured and the result was within maximum crimped stent profile measurement.The balloon cones were reviewed and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.The tip was visually and microscopically examined and no signs of damage were noted.A visual and tactile examination found multiple kinks along several locations of the hypotube shaft.A visual and tactile examination of the outer and mid-shaft section found no issues with the extrusion shaft.The bi-component bond showed no signs of damage or strain.No other issues were identified during the product analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion was unable to be determined.(b)(4).
 
Event Description
Reportable based on device analysis completed on (b)(6) 2018.A 3.50 x 20 synergy ii drug-eluting stent was reported to be defective.No other available information.However, returned device analysis revealed proximal stent damage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY II EVEROLIMUS-ELUTING PLATINUM CHROMIUM CORONARY STENT SYSTEM
Type of Device
BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
Manufacturer (Section G)
BOSTON SCIENTIFIC - GALWAY
Manufacturer Contact
sonali arangil
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key7861946
MDR Text Key119789883
Report Number2134265-2018-07433
Device Sequence Number1
Product Code NIQ
UDI-Device Identifier08714729840183
UDI-Public08714729840183
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P150003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial
Report Date 08/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/10/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/19/2018
Device Model NumberH7493926020350
Device Catalogue Number39260-2035
Device Lot Number0020959415
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/25/2018
Date Manufacturer Received08/18/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/19/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-