• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEXCOM, INC. DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEXCOM, INC. DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number MT22719-PNK
Medical Device Problem Code Unexpected Shutdown (4019)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/21/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
Dexcom was made aware on (b)(4) 2018 that on (b)(6) 2018 the receiver ceased to function.No additional event or patient information is available.No product or data was provided for evaluation.The confirmation of the complaint is undetermined.The probable cause could not be determined.
 
Additional Manufacturer Narrative
Com-(b)(4).
 
Event or Problem Description
The receiver was returned for evaluation.An external visual inspection was performed and failed due to an unreadable serial number label.A receiver charge and boot was performed and passed.A receiver download was performed and rtt loss was observed.A functional test was performed and failed lcd testing.The receiver case was opened, and an internal visual inspection was performed and failed due to moisture damage.Confirmation of the complaint was confirmed.The probable cause could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM
Common Device Name
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence dr.
san diego CA 92121
MDR Report Key7861950
Report Number3004753838-2018-105075
Device Sequence Number1966370
Product Code PQF
UDI-Device Identifier10386270000269
UDI-Public10386270000269
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P120005
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMT22719-PNK
Device Catalogue NumberSTK-RF-PNK
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/13/2018
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 08/21/2018
Supplement Date Received by Manufacturer12/03/2018
Initial Report FDA Received Date09/10/2018
Supplement Report FDA Received Date12/28/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age56 YR
-
-