| Catalog Number |
0375542000 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/14/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that there was foreign material in the sterile packaging.
|
| |
|
Additional Manufacturer Narrative
|
|
The product was returned for investigation and the failure mode will be monitored for future reoccurrence.Alleged failure: a red foreign object was found at the tip of the device.A backup device was used.The failure(s) identified in the investigation is consistent with the complaint record.The probable root cause/s could be the flash was not removed during cleaning process when the inner tube was inserted into the housing and scraped off the flash, improper cleaning process, qc incoming, and in-process inspections.In addition, nc was opened to address the issue.
|
| |
|
Event or Problem Description
|
|
It was reported that there was foreign material in the sterile packaging.
|
| |
|