To date the inserter has not been returned for investigational review.It has been in excess of 49 days from the date created and the agency did indicate that the instrument would be returned for evaluation.When the reported even occurred the instrument may have been in service for 2.7 years.This event occurred during surgery near the patient.There was another suitable device available, the incident did not cause a delay in surgery, surgery was completed as intended, there was no risk or adverse event reported and the instrument was inspected prior to surgery and was found to be acceptable.Instruments are subjected to heavy use/misuse/wear and care must be taken to avoid compromising their performance.A review of the device history record (dhr) revealed when released for use, met design and manufacturing requirements.A review of the complaint history indicates to date, 4 complaints have been reported against this lost number.Summary of complaints for the part number: 2 - functional, 2 - dull/worn, 1 - seized/galled, 44 - thread damaged/galled, 1 - dimensional concern, 3 - bent, 19 - broke/cracked/damaged, 1 - device cracked/broke, 1 - end of life, 32- blank.No further evaluation can be made for this event.This customer complaint will be closed.If the device is returned a complaint amendment will be initiated.The root cause for the initial problem cannot be determined with confidence as the instrument was not returned for investigational review.No containment of inventory required, material or design issues cannot be identified.
|