WRIGHTS LANE SYNTHES USA PRODUCTS LLC TI NUT 11MM WIDTH ACROSS FLATS; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DDD
|
Back to Search Results |
|
| Model Number |
498.003 |
| Medical Device Problem Code |
Unintended Movement (3026)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Patient weight not available for reporting.Additional product codes: mnh, mni, kwp, kwq.Complainant part is not expected to be returned for manufacturer review/investigation.Without a lot number the device history records review could not be completed.The investigation could not be completed; no conclusion could be drawn, as no product was received.(b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2018, the patient had a t11-l3 posterior spinal fusion with synthes universal spinal system (uss) (no screws in l1 - pedicle subtraction osteotomy level).According to the surgeon said that the patient had been experiencing increasing pain in the past month.At an unknown time, the l3 screws slid off the rod caudally.On (b)(6) 2018, the surgeon performed the revision surgery.He removed both of the 6.0 x 125mm rods.He then replaced the l3 7.0 x 55mm uss screws with 8.0 x 55mm screws.The existing screws were slightly loose in the bone.He then selected 6.0 x 150mm rods and connected them to the screws.He final tightened all the nuts and successfully completed the procedure.Concomitant devices reported: titanium hard rod (part 498.105, lot number unknown, quantity 2), titanium collar (part 498.010, lot number unknown, quantity 2), titanium side-opening screw (part 498.755, lot number unknown, quantity 2).This report is for one (1) titanium nut.This is report 1 of 1 for (b)(4).
|
| |
|
Event or Problem Description
|
|
It was further reported that there was no patient consequence.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|