Model Number ZA9003
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 06/25/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that za9003 monofocal intraocular lens (iol) had to be explanted from patient's right eye (od) due to posterior dislocation of lens.Moreover, there was anterior vitrectomy and there was suture required.Product is expected to be returned.No additional information provided.
 
Additional Manufacturer Narrative
Device available for evaluation? yes.Returned to manufacturer on: 8/31/2018.Device returned to manufacturer? yes.Device evaluation: the complaint lens was received.The lens was observed under microscope and it was observed with one haptics positioned and the other slightly bent and no defects and or scratches were observed in the optic zone.The condition observed could be related to the lens removal.The complaint could not be verified in the returned sample since it was a surgery situation.No product quality deficiency was identified.Manufacturing record review: based on the manufacturing records review and related documents the production order was manufactured according specifications.A complaint history review revealed that no other complaint have been received for this production order number.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the product.Conclusion: as a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are (b)(4) and capa-(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS
Common Device Name
MONOFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
santa ana CA 92705
MDR Report Key7862776
Report Number2648035-2018-01224
Device Sequence Number1967051
Product Code HQL
UDI-Device Identifier05050474528925
UDI-Public(01)05050474528925(17)211118
Combination Product (Y/N)Y
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
P990080
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/03/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/18/2021
Device Model NumberZA9003
Device Catalogue NumberZA90030155
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/31/2018
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 08/17/2018
Supplement Date Received by Manufacturer09/28/2018
10/25/2020
Initial Report FDA Received Date09/10/2018
Supplement Report FDA Received Date10/26/2018
11/03/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age74 YR
Date Report Sent to FDA01/01/2005
-
-