| Model Number |
MMT-523NAS |
| Medical Device Problem Code |
No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
Hyperglycemia (1905)
|
| Date of Event |
08/02/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Customer reported via phone call that they experienced high blood glucose.The customer blood glucose level was over 400 mg/dl at the time of incident and the current blood glucose of the patient is unknown.Customer treated with insulin pump and manual injection.Troubleshooting was not performed.The product will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
No unexpected no delivery alarm noted during testing.Device passed rewind test, basic occlusion test, occlusion test, prime test, excessive no delivery test and displacement test.(b)(4).
|
| |
|