• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NEUROSURGERY BECKER EXTERNAL DRAINAGE AND MONITORING SYSTEM; DEVICE, MONITORING, INTRACRANIAL PRESSURE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NEUROSURGERY BECKER EXTERNAL DRAINAGE AND MONITORING SYSTEM; DEVICE, MONITORING, INTRACRANIAL PRESSURE Back to Search Results
Model Number 46128
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 08/21/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported the patient connection line was damaged during the procedure.The cause of the damage was unknown.There was no injury to the patient.
 
Additional Manufacturer Narrative
The device did not meet the requirements for leak testing or patency due to the disconnected tubing and tape that was around the tubing and drainage bag.Therefore, the complaint could not be duplicated.It is unknown how or when this damage occurred.All edms devices are 100% leak tested at the time of manufacture.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
The device did not meet the requirements for leak testing or patency due to the disconnected/damaged tubing.Tubing was damaged and had been taped to the device.It is unknown how or when this damage occurred.All edms devices are 100% leak tested at the time of manufacture.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BECKER EXTERNAL DRAINAGE AND MONITORING SYSTEM
Common Device Name
DEVICE, MONITORING, INTRACRANIAL PRESSURE
Manufacturer (Section D)
MEDTRONIC NEUROSURGERY
125 cremona drive
goleta CA 93117
MDR Report Key7863565
Report Number2021898-2018-00447
Device Sequence Number16478098
Product Code GWM
UDI-Device Identifier00673978228754
UDI-Public00673978228754
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K984053
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/15/2023
Device Model Number46128
Device Catalogue Number46128
Device Lot Number214862975
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/25/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/21/2018
Supplement Date Received by Manufacturer10/15/2018
08/12/2019
Initial Report FDA Received Date09/11/2018
Supplement Report FDA Received Date11/06/2018
08/12/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-