| Model Number |
46128 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/21/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported the patient connection line was damaged during the procedure.The cause of the damage was unknown.There was no injury to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The device did not meet the requirements for leak testing or patency due to the disconnected tubing and tape that was around the tubing and drainage bag.Therefore, the complaint could not be duplicated.It is unknown how or when this damage occurred.All edms devices are 100% leak tested at the time of manufacture.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
The device did not meet the requirements for leak testing or patency due to the disconnected/damaged tubing.Tubing was damaged and had been taped to the device.It is unknown how or when this damage occurred.All edms devices are 100% leak tested at the time of manufacture.If information is provided in the future, a supplemental report will be issued.
|
| |
|