• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS SYMFONY TORIC; MULTIFOCAL IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS SYMFONY TORIC; MULTIFOCAL IOLS Back to Search Results
Model Number ZXT150
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Date of event: unknown, not provided, but the best estimate date is between 3/6/18 and 08/21/2018.(b)(4).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that a zxt150 12.5 diopter intraocular lens was implanted in the patient's left eye (os) on (b)(6) 2018.It was later explanted on (b)(6) 2018 because the patient needed better near vision.There was no incision enlargement, no vitrectomy, and no sutures used.The replacement lens was the same diopter, but different model zmb00.Reportedly, one-day post-op the patient reported no complications.No additional information was provided.
 
Additional Manufacturer Narrative
Device available for evaluation: yes.Returned to manufacturer on: 10/02/2018.Device returned to manufacturer: yes.Device evaluation: the product was returned to the manufacturing site for evaluation.The product is inspected by a qualified inspector using a microscope with a 12x magnification.It can be seen that the lens is contaminated, dust particles are present.This is expected as the lens was transported from controlled environment during manufacturing to a non-sterile, non-environmental controlled area.Also it can be seen that the optic body has a scratch on the surface and it is cut in.Visual inspection of the return sample shows the product is damaged, most probably to make explant possible.Investigation of the return sample and the condition of the return sample do not suggest the damage was introduced during manufacturing.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search on complaints revealed that no similar complaints were received for this production order number.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SYMFONY TORIC
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
santa ana CA
MDR Report Key7863718
Report Number9614546-2018-00881
Device Sequence Number1911211
Product Code POE
UDI-Device Identifier05050474605053
UDI-Public(01)05050474605053(17)220628
Combination Product (Y/N)Y
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/03/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/28/2022
Device Model NumberZXT150
Device Catalogue NumberZXT150U125
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/02/2018
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 08/17/2018
Supplement Date Received by Manufacturer10/16/2018
10/25/2020
Initial Report FDA Received Date09/11/2018
Supplement Report FDA Received Date10/23/2018
11/03/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Date Report Sent to FDA01/01/2005
-
-