• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. SUTR ANCHBIO-COMP S-TAK KNOTLESS; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. SUTR ANCHBIO-COMP S-TAK KNOTLESS; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE Back to Search Results
Model Number SUTR ANCHBIO-COMP S-TAK KNOTLESS
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 08/23/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was requested/is expected but has not returned for evaluation.The cause of the event could not be determined from the information available and without device evaluation.
 
Event or Problem Description
It was reported that during a right shoulder arthroscopy labral repair the implant within the ar-1938bc suture anchor, biocomposite suturetak split and was still lodged partially in the instrument.The rn sated that the suture also broke.The ar-1938bc broke and split lengthwise after the implant was seated.The implant remains partially in the patient, and no other implants were used.Additional information has been requested.Additional information received on 09/07/2018: the facility confirmed that half of the implant was left in the patient.The facility confirmed that the device will be returning for evaluation.
 
Additional Manufacturer Narrative
Complaint confirmed.Visual evaluation of the returned ar-1938bc revealed that approximately 1.61mm of the 3.0mm biocomposite anchor at the distal tip of the device remained intact, while the rest had broken off prior to receipt for investigation.The breakage point was noted to be uneven, with the majority of the remainder of the anchor mass at one side of the device.This event is most likely caused by not inserting the implant co-axial to the bone tunnel, or improper bone preparation prior to insertion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUTR ANCHBIO-COMP S-TAK KNOTLESS
Common Device Name
FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
MDR Report Key7863804
Report Number1220246-2018-00639
Device Sequence Number16535975
Product Code MAI
UDI-Device Identifier00888867097186
UDI-Public00888867097186
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
K180594
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/30/2020
Device Model NumberSUTR ANCHBIO-COMP S-TAK KNOTLESS
Device Catalogue NumberAR-1938BC
Device Lot Number10212559
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/19/2018
Initial Date Received by Manufacturer 08/28/2018
Supplement Date Received by Manufacturer08/28/2018
Initial Report FDA Received Date09/11/2018
Supplement Report FDA Received Date11/29/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age34 YR
Patient Weight53
-
-