• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JOHNSON AND JOHNSON SURGICAL VISION, INC. TECNIS ITEC PRELOADED 1-PIECE IOL; MONOFOCAL IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JOHNSON AND JOHNSON SURGICAL VISION, INC. TECNIS ITEC PRELOADED 1-PIECE IOL; MONOFOCAL IOLS Back to Search Results
Model Number PCB00
Medical Device Problem Codes Material Separation (1562); Difficult to Advance (2920); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/23/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If implanted, give date: not applicable as the lens was not implanted.If explanted, give date: not applicable as the lens was not implanted and therefore not explanted.(b)(6).All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that as the intraocular lens (iol) was being inserted into the patient's operative eye, the trailing haptic got stuck in the injector.Reportedly, the surgeon tried to release the haptic, but it broke off and it appeared to be still in the injector.The lens had to be removed from the eye and another lens was implanted.There was no anterior vitrectomy or suture required.There was a 10-minute delay in the procedure while the lens was removed.No further information was provided.
 
Additional Manufacturer Narrative
Device available for evaluation? yes.Returned to manufacturer on: 11/12/2018.Device returned to manufacturer? yes.Device evaluation: the preloaded delivery system, model pcb00, was returned at the manufacturing site for evaluation.Visual inspection under the microscope showed scarce amount of viscoelastic in the cartridge.Dfu (direction for use) states to completely fill the viewing window of the pcb00 with ovd (ophthalmic viscosurgical device).The haptic was observed detached, stuck in the cartridge and overridden by the plunger.No assembly error and/or defect were observed in the preloaded device related to the manufacturing process.The condition of the returned sample was consistent with a product that was handled and prepared for a surgical process.The complaint issue as stuck in cartridge, override and detached haptic was confirmed.Due to the condition of the returned pcb00 tecnis 1- piece model, the complaint could not be related to manufacturing process, but it could be related to handling process.Therefore, a product quality deficiency was not identified.Manufacturing record review: the manufacturing process record was evaluated and no deviation was found during process related to the complaint issue reported.There was no discrepancy found during the mrr (manufacturing record review).The product was manufactured and released according to specifications.A search revealed that no other complaint was received from this production order.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the product.Conclusion: as a result of the investigation there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS ITEC PRELOADED 1-PIECE IOL
Common Device Name
MONOFOCAL IOLS
Manufacturer (Section D)
JOHNSON AND JOHNSON SURGICAL VISION, INC.
santa ana CA
MDR Report Key7863807
Report Number2648035-2018-01226
Device Sequence Number1703469
Product Code HQL
UDI-Device Identifier05050474558205
UDI-Public(01)05050474558205(17)201117
Combination Product (Y/N)Y
Initial Reporter CountryUK
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Nurse
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/03/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/17/2020
Device Model NumberPCB00
Device Catalogue NumberPCB0000170
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/12/2018
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 08/24/2018
Supplement Date Received by Manufacturer11/19/2018
10/25/2020
Initial Report FDA Received Date09/11/2018
Supplement Report FDA Received Date12/10/2018
11/03/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA01/01/2005
-
-