| Model Number |
9-ITV12F45/80 |
| Medical Device Problem Code |
Material Perforation (2205)
|
| Health Effect - Clinical Codes |
Death (1802); Pericardial Effusion (3271)
|
| Date of Event |
08/24/2018
|
|
Type of Reportable Event
|
Death
|
|
Additional Manufacturer Narrative
|
|
An event of cardiac perforation, effusion and patient death was reported.The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
| |
|
Event or Problem Description
|
|
This case was a "surgical turn down" due to the condition of the patient.During a procedure to close a post infarct ventricular septal defect (vsd), the defect was crossed using the standard method left to right technique to create an av loop.The device access was via transjugular approach.The 12f torqvue 45 delivery system (dtv45) was introduced and crossed through the defect to the right ventricle and then to the left ventricle over the guidewire.A 24 mm amplatzer p.I.Muscular vsd occluder (pimuscvsd) (lot: 6351799) was selected and deployed in an inferior apical defect.The left ventricular disc prolapsed through the defect without applying tension as the defect was large so the pimuscvsd device was removed and a 30 mm amplatzer septal occluder (aso) was then attempted to be used off-label.The dtv45 sheath was still in the right ventricle, however; the guidewire was no longer across the defect due to breaking the circuit to allow deployment of the 24mm pimuscvsd.The user decided that due to the size of the defect they would try and cross from the right ventricle to the left ventricle using a pigtail catheter and guidewire.The defect was crossed and the pigtail catheter was positioned in the aorta for stability.The echo cardiologist noticed that the dtv45 sheath had crossed the defect.The angle was different to the first position (sitting at the apex of the ventricle) and there was concern the user had gone through another defect, so sometime was spent checking the position of the dtv45 to confirm it was through the same defect.The left disc of the 30mm aso was deployed in the right ventricle, but at this time an effusion was noted on echocardiogram by the cardiologist.The aso was fully recaptured and the 12f amplatzer torqvue 45 delivery system was pulled back into the right side of the heart.The effusion was noted to have been caused by rupture of the free wall and the patient died during the procedure.The cause of the rupture of the free wall is unknown.
|
| |
|
Event or Problem Description
|
|
This case was a "surgical turn down" due to the condition of the patient.During a procedure to close a post infarct ventricular septal defect (vsd), the defect was crossed using the standard method left to right technique to create an av loop.The device access was via transjugular approach.The 12f torqvue 45 delivery system (dtv45) was introduced and crossed through the defect to the right ventricle and then to the left ventricle over the guidewire.A 24 mm amplatzer p.I.Muscular vsd occluder (pimuscvsd) (lot: 6351799) was selected and deployed in an inferior apical defect.The left ventricular disc prolapsed through the defect without applying tension as the defect was large so the pimuscvsd device was removed and a 30 mm amplatzer septal occluder (aso) was then attempted to be used off-label.The dtv45 sheath was still in the right ventricle, however; the guidewire was no longer across the defect due to breaking the circuit to allow deployment of the 24mm pimuscvsd.The user decided that due to the size of the defect they would try and cross from the right ventricle to the left ventricle using a pigtail catheter and guidewire.The defect was crossed and the pigtail catheter was positioned in the aorta for stability.The echo cardiologist noticed that the dtv45 sheath had crossed the defect.The angle was different to the first position (sitting at the apex of the ventricle) and there was concern the user had gone through another defect, so sometime was spent checking the position of the dtv45 to confirm it was through the same defect.The left disc of the 30mm aso was deployed in the left ventricle, but at this time an effusion was noted on echocardiogram by the cardiologist.The aso was fully recaptured and the 12f amplatzer torqvue 45 delivery system was pulled back into the right side of the heart.The effusion was noted to have been caused by rupture of the free wall and the patient died during the procedure.The cause of the rupture of the free wall is unknown.Medical records, including echocardiogram reports, were requested from the hospital but have not been made available.
|
| |
|
Search Alerts/Recalls
|