Model Number 179704880
Medical Device Problem Code Break (1069)
Health Effect - Clinical Codes Spinal Column Injury (2081); Not Applicable (3189)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).A complaint investigation will be performed.The complaint product is not available for the investigation.A supplemental report is not anticipated unless the results of the complaint investigation identify a corrective action or additional relevant information.Should the product become available, a physical evaluation will be conducted and a supplemental report filed with the results.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the primary surgery was performed on (b)(6) 2015 by using expedium plus spine system for sacroiliac joint.On (b)(6) 2018, the patient complained the pain at sacroiliac joint, so that the surgeon found that the reported screw was broken under x-ray (it was unknown when the breakage was occurred).It was reported that the only broken screw was explanted on (b)(6) 2018 from the patient¿s sacroiliac joint.No further information was provided by the hospital.
 
Additional Manufacturer Narrative
(b)(4).Returned device was sent for fracture analysis.The screw fracture surfaces image does not suggest fatigue fracture as characteristic fatigue striations are absent.The surfaces show highly irregularly shaped faces from either ductile failure and/or in vivo fretting.The bottom part of the shank displays apparent damage to the fracture surface and external threads consistent with tool marks inflicted during shank removal post fracture.A review of the device history record was conducted.No issues were identified during the manufacturing and release of this product that could have contributed to the problem reported by the customer.All complaint trends will be evaluated as a part of the depuy spine monthly complaint review meeting.A definitive root cause for the post-operative breaking of the reported poly-axial screw cannot be determined.Fracture analysis of the screw suggests that surfaces show highly irregularly shaped faces from either ductile failure and/or in vivo fretting.Surfaces also display apparent damage to the fracture surface and external threads consistent with tool marks inflicted during shank removal post fracture.As there has been no issue identified in the manufacturing or release of the device that could have contributed to the problem reported by the customer and no systemic trends were found, this complaint file will be closed with no further action required.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIPER TI SAI POLY 8X80MM
Common Device Name
ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
MDR Report Key7864024
Report Number1526439-2018-50858
Device Sequence Number1720249
Product Code NKB
UDI-Device Identifier10705034265728
UDI-Public(01)10705034265728
Combination Product (Y/N)N
PMA/510(K) Number
K111571
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2018
Reporter Type Manufacturer
Report Source foreign,user facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number179704880
Device Catalogue Number179704880
Device Lot NumberTBIXT
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/17/2018
Initial Date Received by Manufacturer 08/13/2018
Supplement Date Received by Manufacturer09/20/2018
Initial Report FDA Received Date09/11/2018
Supplement Report FDA Received Date09/20/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age62 YR
-
-