| Model Number |
X SERIES |
| Medical Device Problem Code |
Defibrillation/Stimulation Problem (1573)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has not received the product for evaluation and this complaint is still under investigation.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that the during a routine shift check by a clinician, the device displayed a "defib disabled" message.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical singapore for evaluation; the device's activity log was forwarded to zoll medical us.Review of the device's activity log does show occurrences of a defib failure prompt.The defib board was replaced as a precaution and will be forwarded to zoll medical us for further evaluation.Analysis of reports of this type has not identified an increase in trend.
|
| |
|
Additional Manufacturer Narrative
|
|
The processor/bridge/pace board was returned to zoll medical us; the malfunction was duplicated and attributed to a faulty relay on the processor/bridge/pace.Analysis for reports of this type has not identified an increase in trend.
|
| |
|