C.R. BARD, INC. (COVINGTON) -1018233 BARDIA® COUDE OLIVE TIP URETHRAL FOLEY CATHETER; COUDE CATHETER Back to Search Results
Model Number 802512
Medical Device Problem Code Misassembled (1398)
Health Effect - Clinical Codes No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device was not returned.
 
Event or Problem Description
It was reported that the marked coude catheters seemed straight.No medical intervention was required.
 
Additional Manufacturer Narrative
The reported event was unconfirmed.130 unopened coude olive tip urethral catheters were received for evaluation.The visual evaluation of the devices noted that they were unused and in good condition.The samples were size tested.The catheters were observed to be not straight.No burrs, tackiness, roughness, sharp lumps, or smooth lumps were found.All measured to be 12 fr.Twenty nine samples were evaluated and all were curved and met specifications.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "indications for use: for urological use only.Storage: store catheters at room temperature away from direct exposure to light, preferably in the original box.Caution: this product contains natural rubber latex which may cause allergic reactions.Warnings: ¿ on latex and red rubber catheters, do not use ointments or lubricants having a petrolatum base.They will damage latex.¿ visually inspect the product for any imperfections or surface deterioration prior to use.¿ reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to failure, and/or to injury, illness or death of the patient.Note: after use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practice and applicable local, state and federal laws and regulations.Adverse reactions urinary tract infection bleeding from the urethra irritation of the urethra instructions for intermittent catheters: 1.Wash your hands thoroughly with soap and water.2.Remove catheter from the pack.3.Position yourself comfortably, cleaning the opening of the urethra and surrounding area.4.Gently insert rounded end of catheter into urethra.5.When urine stops flowing, remove catheter from urethra.6.Dispose of catheter in accordance with local rules and regulations.7.Wash your hands.Caution: federal (u.S.A.) laws restricts this device to sale by or on the order of a physician." section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that the marked coude catheters seemed straight.No medical intervention was required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDIA® COUDE OLIVE TIP URETHRAL FOLEY CATHETER
Common Device Name
COUDE CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key7864356
Report Number1018233-2018-04099
Device Sequence Number19317406
Product Code KOD
UDI-Device Identifier00801741039034
UDI-Public(01)00801741039034
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,distributor,other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 02/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date06/30/2019
Device Model Number802512
Device Catalogue Number802512
Device Lot NumberNGBT1127
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/12/2018
Initial Date Received by Manufacturer 08/21/2018
Supplement Date Received by Manufacturer01/30/2019
Initial Report FDA Received Date09/11/2018
Supplement Report FDA Received Date02/21/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-