If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The complaint device was received and evaluated.Visual observation shows the shaft of the device is bent in the middle.The shrink tubing near the distal end of the electrode is damaged possibly by coming in contact with a sharp metal surface.This type of damage indicates rough handling.There were no signs of melting on the shrink tubing.The device was tested for the articulation feature and it functioned as intended.This complaint cannot be confirmed.A review into the complaints system revealed no other complaints for this lot of devices that was released to distribution.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.A review of the device history record indicated that this batch of product was processed without incident; therefore, there is no evidence of manufacturing anomalies on the paperwork reviewed.Udi: (b)(4) - incomplete.The lot number is currently unavailable; therefore, the device expiration date is also unavailable.This report is being filed from the (b)(4) as required under mitek's corrective and preventative actions (capa) to file usa fda mdr missed malfunctions.
|