It was reported, an advance 35 lp low profile balloon catheter ruptured circumferentially after the second inflation while treating a heavily calcified lesion of the superficial femoral artery.The vessel itself was not calcified or tortuous.A 50/50 mix of visipaque to saline solution was in use at the time.The complaint device did not reach nominal pressure, estimated 6 atm (atmospheres) was achieved.According to the complainant this occurred during a lower extremity angiogram procedure.During the procedure, the physician ballooned the popliteal artery with a different cook balloon catheter over another manufacturer's guide wire without issue.The complaint device was then used for a second lesion and, upon inflation, it burst circumferentially, reportedly due to the heavy calcification of that lesion.Removal of the complaint device was said to be difficult and was attempted over another manufacturer's guide wire by pulling it straight out versus using a clockwise rotation as listed in the instructions for use.The device then "bunched up" when it reached the 5fr 55cm cook sheath.As a result, the vessel was perforated causing concern for hematoma formation and all devices were removed.Blood loss was minimal as manual pressure was applied.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.The patient reportedly was fine and was discharged from the facility.
|
|
Investigation ¿ evaluation.A review of the dimensional verification, complaint history, device history record, manufactures instructions, quality control, trends, and a functional test & visual inspection of the returned device were conducted during the investigation.The visual inspection and functional test of the returned device confirmed that the balloon had been returned lodged inside 5 fr.Sheath.The catheter was exposed 86.5cm from the proximal end of the sheath.It was exposed 5.0cm at the distal end of the sheath.A kink was noted 79.5cm from the proximal end of the catheter.Both marker bands were present; however, there were 9.0cm long compression marks noted 96.5cm from the proximal end.Additionally, the balloon separated radially with a proximal section of 3.1cm and a distal section of 6.8cm.A 7.0cm portion of catheter tubing was noted between the two balloon segments.The distal segment of balloon had accordion damage.It was also noted that the full length of the balloon catheter was 148.0cm, which exceeds the specifications for length.This was due to the damage of the catheter tubing.Additionally, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.A review of the device history record showed no nonconforming events which could contribute to this failure mode.It should be noted there were no other reported complaints for this lot number.Furthermore, an ifu is provided with the device, which states ¿if resistance is met while advancing the balloon dilation catheter, determine the cause and proceed with caution.¿ it goes on to say "if balloon pressure is lost and/or balloon rupture occurs, deflate balloon and remove balloon and sheath as a unit." and "upon catheter withdrawal, a gentle counterclockwise rotation of the catheter will assist balloon rewrap, minimizing trauma to the percutaneous entry site." moreover, a review of the manufactures instructions and of quality control revealed no gaps in processing.Based on the information provided, the examination of the returned product, and the results of our investigation, a definitive root cause could not be traced to the device but may lie with the user and patient condition/anatomy.Measures are being conducted to address this failure mode, and we will continue to monitor for similar complaints.Per the quality engineering risk assessment no further action is required.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|