• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC ADVANCE 14 LP LOW PROFILE BALLOON CATHETER; DQY CATHETER, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC ADVANCE 14 LP LOW PROFILE BALLOON CATHETER; DQY CATHETER, PERCUTANEOUS Back to Search Results
Catalog Number PTAX4-14-170-2-4
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 08/22/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.
 
Event or Problem Description
It was reported, during an unspecified endovascular therapy (evt) procedure, an advance 14 lp low profile balloon catheter ruptured when it was inflated to nominal pressure in a severely calcified lesion.Brachial approach was used, and another manufacturer's balloon was used to complete the procedure with no adverse effect to the patient.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Additional Manufacturer Narrative
Investigation ¿ evaluation a review of the complaint history, device history record, drawing, instructions for use(ifu), manufactures instructions, quality control, trends, and specifications of the device was conducted during the investigation.The complaint device was not returned; therefore, no physical examinations could be performed.However, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.A review of the device history record shows no nonconforming events which could contribute to this failure mode.There were two other reported complaints with this lot number.However, it should be noted that they are unrelated to this failure mode.Furthermore, a review of quality control and manufactures instructions revealed no gaps.Moreover, an ifu is provided with the device, which states ¿the balloon is manufactured from an extra-thinwall, high-strength, minimally-compliant material.Particular care should be taken in handling the balloon to prevent damage.It will inflate to the indicated size parameters when utilizing proper pressure recommendations.Adhere to balloon inflation pressure parameters indicated in the compliance card insert.¿ based on the information provided, no product returned and the results of our investigation, a definitive root cause cannot be traced to the device, and likely lies with the patient¿s anatomy (calcification).We will continue to monitor for similar complaints.Per the quality engineering risk assessment no further action is required.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCE 14 LP LOW PROFILE BALLOON CATHETER
Common Device Name
DQY CATHETER, PERCUTANEOUS
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7865491
Report Number1820334-2018-02645
Device Sequence Number19252134
Product Code DQY
UDI-Device Identifier10827002503141
UDI-Public(01)10827002503141(17)190624(10)7075376
Combination Product (Y/N)N
PMA/510(K) Number
K090822
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/24/2019
Device Catalogue NumberPTAX4-14-170-2-4
Device Lot Number7075376
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/22/2018
Supplement Date Received by Manufacturer10/26/2018
Initial Report FDA Received Date09/11/2018
Supplement Report FDA Received Date11/02/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
BOSTON SCIENTIFIC COYOTE CATHETER
-
-