MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) CURRENT DR RF; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) CURRENT DR RF; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number 2207-36
Medical Device Problem Code Device Sensing Problem (2917)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 08/28/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event or Problem Description
It was reported that patient presented for a device change out due to a variable r-wave sensing that had been noted prior.The device was explanted and replaced successfully.The patient was stable prior during and after the procedure.
 
Additional Manufacturer Narrative
Bench testing was performed, which includes impedance, sensing, pacing, and high voltage output tests and the device was normal.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CURRENT DR RF
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
MDR Report Key7865642
Report Number2017865-2018-13357
Device Sequence Number19224172
Product Code LWS
UDI-Device Identifier05414734501927
UDI-Public05414734501927
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
P910023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2010
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2010
Device Model Number2207-36
Device Lot Number3096378
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/11/2018
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 08/28/2018
Supplement Date Received by Manufacturer03/18/2019
Initial Report FDA Received Date09/11/2018
Supplement Report FDA Received Date03/21/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age82 YR
-
-