MAUDE Adverse Event Report: MSD DEGGENDORF MFG PEEK PREVAIL CERVICAL INTERBODY DEVICE; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, CERVICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MSD DEGGENDORF MFG PEEK PREVAIL CERVICAL INTERBODY DEVICE; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, CERVICAL Back to Search Results
Catalog Number 4210564
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 08/20/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Type of procedure or technique used: anterior cervical discectomy pre-operative diagnosis: cervical spine instability it was reported that intra-op, implant broke during the implantation.The product came in contact with the patient.No patient symptoms or complications were reported in this event.
 
Additional Manufacturer Narrative
Product analysis: visual review confirmed the implant is broken at the base of the vertebral spacer portion of the implant, with asymmetrical deformation at the top and bottom of the face of the implant, consistent with significant impaction forces during attempted implantation.Microscopic examination of the fracture identified a fairly brittle fracture surface, with rays emanating from the origin of crack propagation.The above observations are consistent with overload during attempted insertion of the implant.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PEEK PREVAIL CERVICAL INTERBODY DEVICE
Common Device Name
INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, CERVICAL
Manufacturer (Section D)
MSD DEGGENDORF MFG
wertstrasse 17
deggendorf 94469
MDR Report Key7865839
Report Number1030489-2018-01243
Device Sequence Number1703446
Product Code ODP
UDI-Device Identifier00613994502087
UDI-Public00613994502087
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K073285
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/18/2025
Device Catalogue Number4210564
Device Lot Number84EX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/10/2018
Initial Date Received by Manufacturer 08/20/2018
Supplement Date Received by Manufacturer12/14/2018
Initial Report FDA Received Date09/11/2018
Supplement Report FDA Received Date12/27/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-