• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC 4.5MM VA-LCP CURVED CONDYLAR PLATE/16 HOLE/336MM/LEFT; IMPLANT, FIXATION DEVICE, CONDYLAR PLATE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC 4.5MM VA-LCP CURVED CONDYLAR PLATE/16 HOLE/336MM/LEFT; IMPLANT, FIXATION DEVICE, CONDYLAR PLATE Back to Search Results
Model Number 02.124.417
Medical Device Problem Code Break (1069)
Health Effect - Clinical Codes Bone Fracture(s) (1870); Impaired Healing (2378); No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional product codes: hrs, hwc.Lot number unavailable.Implant and explant dates unknown.Complainant part is not expected to be returned for manufacturer review/investigation.The investigation could not be completed; no conclusion could be drawn, as no product was received.(b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on an unknown date, the patient underwent a revision surgery due to a non-union fracture in the distal femur.The variable angle (va)-locking compression plate (lcp) curved condylar plate broke at the fracture site and was removed.A new plate was placed along the distal femur along with a retrograde femoral nail in canal.It is unknown if there was a surgical delay.Patient outcome is unknown.Concomitant device/s reported: unknown locking screws (part# unknown, lot# unknown, quantity# unknown).This is report 1 of 1 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4.5MM VA-LCP CURVED CONDYLAR PLATE/16 HOLE/336MM/LEFT
Common Device Name
IMPLANT, FIXATION DEVICE, CONDYLAR PLATE
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key7866029
Report Number2939274-2018-53701
Device Sequence Number1678743
Product Code JDP
UDI-Device Identifier10886982042958
UDI-Public(01)10886982042958
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K110354
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number02.124.417
Device Catalogue Number02.124.417
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/21/2018
Initial Report FDA Received Date09/11/2018
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age74 YR
Patient Weight87
-
-