| Model Number |
1430US |
| Medical Device Problem Code |
Power Problem (3010)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/29/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This device is used for treatment not diagnosis.The ventricular assist system is indicated for use as a bridge to cardiac transplantation and destination therapy in patients who are at risk of death from refractory end-stage left ventricular heart failure.The system is designed for in-hospital and out-of-hospital settings.Other devices involved in this event: heartware ventricular assist system ¿ controller ac adapter/ (b)(4)/ model #: 1430us / return date: 2018-09-10 device evaluation anticipated, but not yet begun (b)(4).If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that multiple controller ac adaptors would not provide power when connected to the controller.The multiple adaptors were exchanged.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
The controller was returned for evaluation.Various analyses were conducted and reviewed in order to evaluate the performance of the device in relation to the reported event.Failure analysis of the returned controller revealed bent pins within power port one (1) of the controller.The bent pins did not allow a battery to properly connect to the controller.As a result, the reported event was confirmed.The most likely root cause of the reported event can be attributed to a misalignment between the power port and battery output connector.In internal investigation was opened to investigate bent/damaged pins with controller 2.0.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|