MAUDE Adverse Event Report: HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - CONTROLLER AC ADAPTER; VENTRICULAR (ASSISST) BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - CONTROLLER AC ADAPTER; VENTRICULAR (ASSISST) BYPASS Back to Search Results
Model Number 1430US
Medical Device Problem Code Power Problem (3010)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 08/29/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This device is used for treatment not diagnosis.The ventricular assist system is indicated for use as a bridge to cardiac transplantation and destination therapy in patients who are at risk of death from refractory end-stage left ventricular heart failure.The system is designed for in-hospital and out-of-hospital settings.Other devices involved in this event: heartware ventricular assist system ¿ controller ac adapter/ (b)(4)/ model #: 1430us / return date: 2018-09-10 device evaluation anticipated, but not yet begun (b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that multiple controller ac adaptors would not provide power when connected to the controller.The multiple adaptors were exchanged.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
The controller was returned for evaluation.Various analyses were conducted and reviewed in order to evaluate the performance of the device in relation to the reported event.Failure analysis of the returned controller revealed bent pins within power port one (1) of the controller.The bent pins did not allow a battery to properly connect to the controller.As a result, the reported event was confirmed.The most likely root cause of the reported event can be attributed to a misalignment between the power port and battery output connector.In internal investigation was opened to investigate bent/damaged pins with controller 2.0.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTWARE VENTRICULAR ASSIST SYSTEM - CONTROLLER AC ADAPTER
Common Device Name
VENTRICULAR (ASSISST) BYPASS
Manufacturer (Section D)
HEARTWARE, INC.
14400 nw 60th ave
miami lakes FL 33014
MDR Report Key7866388
Report Number3007042319-2018-04357
Device Sequence Number19258255
Product Code DSQ
UDI-Device Identifier00888707002738
UDI-Public00888707002738
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P100047
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 12/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date01/19/2019
Device Model Number1430US
Device Catalogue Number1430US
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/10/2018
Initial Date Received by Manufacturer 08/29/2018
Supplement Date Received by Manufacturer12/11/2018
Initial Report FDA Received Date09/11/2018
Supplement Report FDA Received Date12/13/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1538-2017
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MCS VAD
Patient Age37 YR
-
-