| Model Number |
10015612 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/24/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Concomitant medical products: 8110, 8015, bd 60 ml syringe (filled to 50 ml); icu medical microclave; therapy date: (b)(6) 2018.Although requested, the affected product has not been received.A follow up report will be submitted with investigation results should the affected product be received for evaluation.
|
| |
|
Event or Problem Description
|
|
The customer reported that during a heparin infusion, the leak was discovered with the last 10-15 ml of the infusion between the needleless valve and the psd tubing.The rate was under 5 ml/hr.The clinician obtained a new syringe and tubing to continue the patient¿s infusion.No patient harm was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The affected product has been received and the evaluation is pending.A follow up report will be submitted once the evaluation is completed.
|
| |
|
Event or Problem Description
|
|
The customer reported that during a heparin (50 units in 50 ml sodium chloride 0.9%) intraarterial infusion, the leak was discovered with the last 10-15 ml of the infusion between the needleless valve and the psd tubing.The rate was under 5 ml/hr.The clinician obtained a new syringe and tubing to continue the patient¿s infusion.No patient harm was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer¿s report of a leak was not confirmed.Visual inspection observed no anomalies.Functional and pressured testing was performed.No issues were observed.The root cause of the reported failure was not identified.
|
| |
|
Event or Problem Description
|
|
The customer reported that during a heparin (50 units in 50 ml sodium chloride 0.9%) intraarterial infusion, the leak was discovered with the last 10-15 ml of the infusion between the needleless valve and the psd tubing.The rate was under 1 ml/hr.The clinician obtained a new syringe and tubing to continue the patient¿s infusion.No patient harm was reported.
|
| |
|