| Model Number |
45-751632 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/13/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The lot number was provided.A review of the approved device history record indicated the lot met all release criteria.A lot history trending review was performed and there were no similar complaints for this lot number.The device has not been returned to the manufacturer for analysis.
|
| |
|
Event or Problem Description
|
|
It was reported that during the embolization treatment for a splenic artery aneurysm, the coil was not forming to the satisfaction of the physician.During retraction of the coil to reposition it, the implant coil detached in the catheter, with 1/2 of the coil in the catheter and 1/2 of the coil in the artery.Both segments were removed in their entirety together with the catheter from the patient.There was no reported intervention or patient injury.
|
| |
|
Search Alerts/Recalls
|