| Model Number |
C42010E |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Although requested, device has not been received.A follow up report will be submitted with failure investigation results should the device be received for evaluation.
|
| |
|
Event or Problem Description
|
|
The customer reported four events where the iv sets came apart, some during administration.No patient harm was reported.This is the file for one event.
|
| |
|
Event or Problem Description
|
|
The customer reported four events where the iv sets came apart, some during administration.No patient harm was reported.This is the file for one event.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer¿s report of iv sets coming apart was confirmed.Visual inspection showed no anomalies.Gentle tugging of the tubing resulted in a separation of the lower portion of the tubing from the stopcock.The root cause of the failure was determined to be failure to tighten the stopcock connections prior to use.
|
| |
|
Event or Problem Description
|
|
The customer reported four events where the iv sets came apart, some during administration.No patient harm was reported.This is the file for one event.
|
| |
|