MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-551PNAL; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-551PNAL; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-551PNAL
Medical Device Problem Code Power Problem (3010)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/20/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
Customer reported via phone call that the batteries died and insulin pump has a blank display even when batteries are replaced.Trouble shooting occurred to try and fix the problem but issue remains unresolved.Blood glucose level at the time of incident is 110 mg/dl.No symptoms reported and therefore no treatment required.Product will be returned for analysis.
 
Additional Manufacturer Narrative
Device received with blank display due to moisture damage at electronic assembly.Also, moisture damage motor during visual inspection.Unable to perform operating currents, self test, rewind test and displacement test or verify battery out limit due to blank display.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
530G INSULIN PUMP MMT-551PNAL
Common Device Name
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key7867734
Report Number2032227-2018-27008
Device Sequence Number1719748
Product Code OZO
UDI-Device Identifier00643169507289
UDI-Public(01)00643169507289
Combination Product (Y/N)N
Initial Reporter StateDE
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-551PNAL
Device Catalogue NumberMMT-551PNAL
Device Lot NumberA3551PNALJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/29/2018
Initial Date Received by Manufacturer 08/20/2018
Supplement Date Received by Manufacturer12/19/2018
Initial Report FDA Received Date09/12/2018
Supplement Report FDA Received Date12/21/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age71 YR
Patient Weight145
-
-