| Model Number |
MMT-551PNAL |
| Medical Device Problem Code |
Power Problem (3010)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/20/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Customer reported via phone call that the batteries died and insulin pump has a blank display even when batteries are replaced.Trouble shooting occurred to try and fix the problem but issue remains unresolved.Blood glucose level at the time of incident is 110 mg/dl.No symptoms reported and therefore no treatment required.Product will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device received with blank display due to moisture damage at electronic assembly.Also, moisture damage motor during visual inspection.Unable to perform operating currents, self test, rewind test and displacement test or verify battery out limit due to blank display.
|
| |
|