| Catalog Number |
05401470190 |
| Medical Device Problem Code |
Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/21/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This event occurred in (b)(6).(b)(6).
|
| |
|
Event or Problem Description
|
|
The customer received questionable high crep2 creatinine plus ver.2 results for seven patient samples from the cobas c 111 with ise analyzer serial number (b)(4).Refer to the attachment to the medwatch for all patient data.The erroneous results were not reported outside of the laboratory.There was no allegation of an adverse event.Calibration and qc results were acceptable and a general reagent issue could be excluded.The probe was replaced and the water quality was checked.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigations have determined a reagent carryover to the crep2 (creatinine) or trigl (triglycerides) test can occur if a preceding test containing dextran sulfate is pipetted first and then followed by the chol2 (cholesterol gen.2) reagent.This carryover can result in discrepant, high results for crep2 and trigl.Installation of extra wash cycles avoids this issue.Customers have been provided with the workaround.
|
| |
|