| Model Number |
231225 |
| Medical Device Problem Code |
Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|
| |
|
Event or Problem Description
|
|
It was reported that there was plastic coming out of the inside of the sheath.The plastic found inside of the sheath was not part of the device.There was no reported patient impact.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was inconclusive.The sample was received used.Observation noted that the sample sheath was partially inside tytek packaging and had a slight bend.The tytek pouch label read 12/14fr and the returned sample sheath was a 10/12 fr size.There was no dilator present in the bag with the sample.The investigation was inconclusive due to the sample received.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: ¿directions for use: 1.Activate the hydrophilic coating by placing the dilator and sheath components into saline or sterile water.Place an 0.035" (0.889mm) or 0.038" (0.965mm) guidewire into the ureter using standard endourology techniques.2.Ensure the dilator lock is securely engaged with sheath hub prior to insertion.3.Insert the guidewire into the tapered end of the dilator/sheath assembly and gradually advance the assembly into the ureter.Note: placement of the assembly can be verified using fluoroscopy or radiographic means.4.While maintaining sheath position, disengage the dilator lock from the sheath hub to gently remove the dilator.Do not advance sheath without the dilator in place.Note: suture holes are provided on sheath hub for securing externally, if desired.5.An endoscope and/or related instruments can now be used through the ureteral sheath as needed.6.If desired, irrigation can be applied using the luer connector on the dilator.7.Upon completion of the access procedure, gently withdraw the device.8.Discard the device in accordance with hospital procedures and with applicable laws and regulations.For urological use only".
|
| |
|
Event or Problem Description
|
|
It was reported that there was plastic coming out of the inside of the sheath.The plastic found inside of the sheath was not part of the device.There was no reported patient impact.
|
| |
|