| Catalog Number |
TPW10 |
| Medical Device Problem Codes |
Break (1069); Tear, Rip or Hole in Device Packaging (2385)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/12/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The device history records were reviewed and the manufacturing criteria was met prior to the release of this lot.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
|
| |
|
Event or Problem Description
|
|
It was reported a patient underwent a valve exchange procedure on (b)(6) 2018 where temporary pacing wire was used.The needle was passing through the package.Product was not used in the patient.Another thread was used.No adverse patient consequences.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # pc-(b)(4).Date sent to the fda: (b)(4) 2018.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # pc-(b)(4).An unopened sample of product was returned for analysis.During the visual inspection of unopened sample, a pin hole on the overwrap package was observed.The package was opened and it was verified that the position of the needle was incorrect.It is probable that the tip of the needle went through the paper folder and the overwrap package.A hole outside the package was visible.Therefore, the sterile barrier was compromised.The manufacturing records were reviewed and the manufacturing/packaging criteria were met prior to the release of this batch.Per the sample condition, the assignable cause of compromise of packaging integrity is a pin hole in package.
|
| |
|