Model Number SDK4000-FT
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Pneumothorax (2012)
Date of Event 08/16/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device was not returned; however the case recordings from the procedure were received and are currently under evaluation.Pneumothorax is a known short-term complication when a lung biopsy is performed during a transbronchial lung biopsy or ct guided percutaneous biopsy.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, post electromagnetic navigation bronchoscopy (enb) procedure, the patient experienced pneumothorax after procedure had been completed.The patient required a chest tube.
 
Additional Manufacturer Narrative
The locatable guide was discarded by the customer.However, a component of the system, case recordings and logs were returned and evaluated.No unusual findings were observed during the investigation.No findings indicating pneumothorax were observed in procedure log analysis.No software malfunction found.Software operated as expected.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDGE
Common Device Name
SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED
Manufacturer (Section D)
COVIDIEN LP - SUPERDIMENSION INC
161 cheshire lane, suite 100
plymouth MN 55441
MDR Report Key7868838
Report Number3004962788-2018-00047
Device Sequence Number19252053
Product Code JAK
UDI-Device Identifier10884521545557
UDI-Public10884521545557
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K102604
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberSDK4000-FT
Device Catalogue NumberSDK4000-FT
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/16/2018
Supplement Date Received by Manufacturer08/16/2018
Initial Report FDA Received Date09/12/2018
Supplement Report FDA Received Date09/18/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SUPERDIMENSION INREACH SYSTEM; SUPERDIMENSION INREACH SYSTEM
Outcome Attributed to Adverse Event Required Intervention;
Patient Age56 YR
Patient Weight90
-
-