Model Number ZMB00
Medical Device Problem Codes Unstable (1667); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Blurred Vision (2137); No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Date of event is unknown as it was not provided.(b)(4).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that patient had unstable implant, where the capsular integrity of the patient's left eye (os) is no longer viable to keep the lens in position.As a result, zmb00 tecnis multifocal iol (intraocular lens) was explanted from patient''s os.Standard incision was also performed to remove the lens and there will be planned anterior vitrectomy per follow-up.Reportedly, the patient had blurred vision (b)(6) 2018, so the lens was also reposition then.However, patient was doing well post-surgery as there was no patient injury.Product is expected to return.No additional information provided.
 
Additional Manufacturer Narrative
Corrected data: in the initial mdr 2648035-2018-01235, (b)(4).Device evaluation: the product was not returned to the manufacturing site.Therefore, product testing could not be performed and the reported complaint could not be verified.Manufacturing records review: the manufacturing records for the product was reviewed.The product was manufactured and released according to specification.A search on complaints revealed that no other complaints have been received for this production order number.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are (b)(4) and capa-(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS 1 MULTIFOCAL
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
santa ana CA 92705
MDR Report Key7868967
Report Number2648035-2018-01235
Device Sequence Number1263831
Product Code MFK
UDI-Device Identifier05050474535534
UDI-Public(01)05050474535534(17)170704
Combination Product (Y/N)Y
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/03/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/04/2017
Device Model NumberZMB00
Device Catalogue NumberZMB00U0155
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 08/20/2018
Supplement Date Received by Manufacturer10/08/2018
10/25/2020
Initial Report FDA Received Date09/12/2018
Supplement Report FDA Received Date11/02/2018
11/03/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age47 YR
Date Report Sent to FDA01/01/2005
-
-