| Model Number |
ZMB00 |
| Medical Device Problem Codes |
Unstable (1667); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Codes |
Blurred Vision (2137); No Code Available (3191)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Date of event is unknown as it was not provided.(b)(4).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that patient had unstable implant, where the capsular integrity of the patient's left eye (os) is no longer viable to keep the lens in position.As a result, zmb00 tecnis multifocal iol (intraocular lens) was explanted from patient''s os.Standard incision was also performed to remove the lens and there will be planned anterior vitrectomy per follow-up.Reportedly, the patient had blurred vision (b)(6) 2018, so the lens was also reposition then.However, patient was doing well post-surgery as there was no patient injury.Product is expected to return.No additional information provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected data: in the initial mdr 2648035-2018-01235, (b)(4).Device evaluation: the product was not returned to the manufacturing site.Therefore, product testing could not be performed and the reported complaint could not be verified.Manufacturing records review: the manufacturing records for the product was reviewed.The product was manufactured and released according to specification.A search on complaints revealed that no other complaints have been received for this production order number.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction mdr is being submitted to address inaccurate information on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are (b)(4) and capa-(b)(4).
|
| |
|