Catalog Number 701044054
Device Problem Defective Device (2588)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/23/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).A follow-up medwatch will be submitted when additional information becomes available.(b)(4).A field service technician has been sent out for investigation on 2018-09-06 and detected the circuit breaker on the patient heater was tripped.Thus the failure could be confirmed.Since no parts were replaced no further investigation could be performed.Therefore no most probable root cause could be determined.According to service (b)(4) the technician reset the circuit breaker of the heater.No safety or function defects were detected according to the service protocol dated on 2018-09-06.The data is also being handled through a designated maquet cardiopulmonary trending and applicable investigation process.Due to this no further investigation initiations will be completed at this time.Since the reported failure did not contribute to a death or serious injury no corrective action is needed.In addition at this time it cannot be concluded that this is a systemic error.No corrective action is needed.A supplemental medwatch will be submitted after new information has been received.
 
Event Description
Self-test failure report: heater pat.Circuit safety shutdown patient circuit failure occurred prior to use.Therefore the surgery was delayed for 30 minutes.No patient injury.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAQUET HCU 40
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY AG
rastatt
GM 
Manufacturer (Section G)
BERND RAKOW
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt
GM  
Manufacturer Contact
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt 
4972229321
MDR Report Key7869167
MDR Text Key120398120
Report Number8010762-2018-00257
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K130300
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number701044054
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/06/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/23/2018
Initial Date FDA Received09/12/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse