Product complaint #: (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Evaluation statement: the complaint device was received and evaluated.Visual observations revealed that the distal tip of the device was broken and the threads appear to be worn.Therefore, this complaint can be confirmed.The information provided stated that the surgeon was not off axis and the patient had standard bone quality.We cannot discern a definitive root cause for the reported failure, however, if there was a possibility that the bone hole was not prepared properly, this could have caused the failure.Furthermore, a nonconformance search was conducted and no nonconformances have been identified for this part 231822 with lot l926609 combination.At this point, no corrective action is required and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Udi: (b)(4).Associated medwatch report number: 1221934-2018-54635.
|