Model Number G07369
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 08/21/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Pma/510(k) number: pre-amendment.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.
 
Event or Problem Description
It was reported, during the inspection of incoming product, it had been observed that there was a particle inside of sealed primary packaging of the performer introducer.The device never made patient contact.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Investigation ¿ evaluation a review of the complaint history, device history record, drawing, manufacturing instructions, quality control, and specifications of the device were conducted during the investigation.The complaint device was not returned; therefore, no physical examinations could be performed.However, photos were provided by the customer.These photos revealed a particle inside the outer pouch of the device set confirming the complaint.Additionally, a document based investigation evaluation was performed.There is evidence to suggest the product was not made to specifications.A review of the device history record shows no nonconforming events which could contribute to this failure mode.It should be noted there were no other reported complaints for this lot number.Furthermore, a review of the manufactures instructions and quality control procedures was conducted, and no gaps were discovered.Based on the information provided, no product returned and the results of our investigation, a definitive root cause could be traced to a manufacturing deficiency.We will continue our monitoring of similar complaints and have notified the appropriate personnel of this event.Per the quality engineering risk assessment no further action is required.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERFORMER INTRODUCER
Common Device Name
DYB INTRODUCER, CATHETER
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7869863
Report Number1820334-2018-02668
Device Sequence Number1720304
Product Code DYB
UDI-Device Identifier00827002073692
UDI-Public(01)00827002073692(17)210327(10)8654422
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/27/2021
Device Model NumberG07369
Device Catalogue NumberRCF-12.0-38-J
Device Lot Number8654422
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/23/2018
Supplement Date Received by Manufacturer12/27/2018
Initial Report FDA Received Date09/12/2018
Supplement Report FDA Received Date01/08/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-